pH/ISE and EC/TDS Benchtop Meter - HI5521-02

Save $103.89
$2,077.90
$1,974.01

Order now, stock will leave our store in the next 2-4 weeks

Email The Electric Burst Team For More Availability Information
SKU HI5521-02
Features: 

• Highly Customizable User Interface - The user interface of the HI5522 allows the user to show measurements in various modes: basic measurement with or without GLP information, real-time graphing, and logging data. 
• Capacitive Touch - The HI5522 features sensitive capacitive touch buttons for accurate keystrokes when navigating menus and screens. There are four dedicated keys that are used for routine operations including calibration and switching measurement modes and four virtual keys that change based upon use. 
• Choice of Calibration - Automatic buffer recognition, semiautomatic, and direct manual entry pH calibration options are available for calibrating up to five points, from a selection of eight standard buffers and up to five custom buffers.
• GLP Data - The IC-HI5522 includes a GLP Feature that allows users to view calibration data and calibration expiration information at the touch of a key. Calibration data include date, time, buffers /standards used for calibration and slope characteristics. The offset is also displayed for pH electrodes.
• CAL Check- CAL Check alerts users to potential problems during the calibration of the pH electrode. Indicators include -Electrode Dirty/Broken,-Buffer Contaminated-, electrode response time and the overall probe condition as a percentage that is based on the offset and slope characteristics.
• Data Logging - Three selectable logging modes are available on the HI5522: automatic, manual, and AutoHold logging. Automatic and manual logs up to 100 lots with 50,000 records max/lot, with up to 100,000 total data points. Automatic logging features the option to save data according to sampling period and interval.

Specifications

pH:

• pH Range: -2.000 to 20.000 pH
• pH Resolution: 0.1, 0.01, 0.001 pH
• pH Accuracy (@25ºC/77ºF): ±0.1 pH, ±0.01 pH, ±0.002 pH ±1 LSD
• pH Calibration: automatic, up to five point calibration, eight standard buffers available (1.68, 3.00, 4.01, 6.86, 7.01, 9.18, 10.01,12.45) and five custom buffers
• pH Temperature Compensation: automatic or manual from -20.0 to 120.0 °C

mV:

• mV Range: ±2000 mV
• mV Resolution: 0.1 mV
• mV Accuracy: ±0.2 mV ±1 LSD
• Relative mV Offset Range: ±2000 mV

EC:

• EC Range: 0.000 to 9.999 µS/cm, 10.00 to 99.99 µS/cm, 100.0 to 999.9 µS/cm, 1.000 to 9.999 mS/cm, 10.00 to 99.99 mS/cm, 100.0 to 1000.0 mS/cm actual EC*
• EC Resolution: 0.001 µS/cm, 0.01 µS/cm, 0.1 µS/cm, 1 µS/cm, 0.001 mS/cm, 0.01 mS/cm, 0.1 mS/cm
• EC Accuracy (@25ºC/77ºF): ±1% of reading (±0.01 µS/cm)
• EC Calibration: automatic standard recognition (0.000 µS/cm, 84.00 µS/cm, 1.413 mS/cm, 5.000 mS/cm, 12.88 mS/cm, 80.00 mS/cm, 111.8 mS/cm) or user standard; single point or multi-point calibration

TDS:

• TDS Range: 0.000 to 9.999 ppm, 10.00 to 99.99 ppm, 100.0 to 999.9 ppm, 1.000 to 9.999 ppt, 10.00 to 99.99 ppt, 100.0 to 400.0 ppt actual TDS* (with 1.00 factor)
• TDS Resolution: 0.001 ppm, 0.01 ppm, 0.1 ppm, 1 ppm, 0.001 ppt, 0.01 ppt, 0.1 ppt
• TDS Accuracy (@25ºC/77ºF): ±1% of reading (±0.01 ppm)
• Resistivity Range: 1.0 to 99.9 O?cm; 100 to 999 O?cm; 1.00 to 9.99 KO?cm; 10.0 to 99.9 KO?cm; 100 to 999 KO?cm; 1.00 to 9.99 MO?cm; 10.0 to 100.0 MO?cm
• Resistivity Resolution: 0.1 O?cm; 1 O?cm; 0.01 KO?cm; 0.1 KO?cm; 1 KO?cm; 0.01 MO?cm; 0.1 MO?cm*
• Resistivity Accuracy: ±2% of reading (±1 O?cm)

Salinity:

• Salinity Range: practical scale: 0.00 to 42.00 psu; natural seawater scale: 0.00 to 80.00 ppt; percent scale: 0.0 to 400.0%
• Salinity Resolution: 0.01 for practical scale/natural seawater scale; 0.1% for percent scale
• Salinity Accuracy (@25ºC/77ºF): ±1% of reading
• Salinity Calibration: percent scale?one-point (with HI7037 standard)

Temperature:

• Temperature Range: -20.0 to 120.0 ºC, -4.0 to 248.0 °F, 253.15 to 393.15 K
• Temperature Resolution: 0.1 °C, 0.1 °F, 0.1 K
• Temperature Accuracy: ±0.2 °C, ±0.4 °F, ±0.2 K
• Temperature Compensation: disabled, linear and non-linear (natural water)
• Electrode/Probe: HI1131B glass body pH electrode with BNC connector and 1 m (3.3?) cable (included)
• Temperature Probe: HI7662-T stainless steel temperature probe with 1 m (3.3?) cable (included)
• GLP: calibration data including date, time, buffers used, offset and slope

ISE:

• ISE Range: 1 x 10?6 to 9.99 x 10?° concentration
• ISE Resolution: 1; 0.1; 0.01; 0.001 concentration
• ISE Accuracy @ 25 °C/77 °F: ±0.5% (monovalent ions); ±1% (divalent ions)
• ISE Calibration Points: automatic, up to five point calibration, five fixed standard solutions available (0.1, 1, 10, 100, 1000 choice of concentration) and five user defined units

You recently viewed

Clear recently viewed