30392104 2 Block Dry Block Heaters With Lid (39.1 X 20.3 X 17.8 cm)

by OHaus
Save $460.84
$3,072.30
$2,611.46

Order now, stock will leave our store in the next 2-4 weeks

Email The Electric Burst Team For More Availability Information
SKU 30392104
The OHAUS Digital Dry Block Heaters with Heated Lid are designed for applications that require repeatable results and superior temperature stability. These multi-purpose units are ideal for isothermal incubation, enzyme reactions, immunoassays, nucleic acid denaturation and a wide variety of other laboratory procedures.

Features:

- Exceptional uniformity, stability, and regulation of temperature
- Heated lid reduces condensation on sample lids
- Optional external temperature probe
- Microprocessor Control: PID temperature control, with optional external RTD probe, offers a temperature stability as low as +/- 0.1 C with a temperature uniformity as low as +/- 0.1 C. Samples are heated to temperature quickly and accurately. Temperature is adjusted in +/- 0.1 C increments.
- LED Display: Touch pad controls with easy-to-read, independent LED displays for temperature, and time. Provides repeatable and accurate results every time and is easily visible across lab benches. Shows set-point and actual temperature.
- Timer will display elapsed time or, when programmed to user defined limit, will shut off unit when time reaches zero. Display will show last used settings, even after power has been turned off.
- Temperature Calibration Mode: Allows user to calibrate unit to an external temperature device.

Safety Features:

- Caution Hot Indicator: Hot warning symbol light is illuminated when the temperature is above 40 C.
- Overshoot Protection: If the unit exceeds the set temperature by 10 C the unit will automatically stop heating.
- Audible Alarm: In timed mode, alarm will sound when time reaches zero and when unit reaches set-point temperature.
- To avoid possible electrical hazard, do not fill well or block opening with water or other fluids. Unit is designed as a dry bath/incubator.

Operating Conditions:

Units can be run in environments from 18 to 33 C (64 to 91 F), 20% to 80% relative humidity, non-condensing.

Applications:

Isothermal incubation, enzyme reactions, immunoassays and nucleic acid denaturation and a wide variety of other laboratory procedures.

Specifications:

- Temperature range: Ambient +5 C to 100 C
- Temperature Stability at 37 C: +/- 0.2 C
- Uniformity Across similar blocks at 37 C: +/- 0.2 C
- Heat-Up Time to 100 C: 50 minutes

You recently viewed